Defaults
The default of the trigger delay is automatic. M3500A automatically
selects a delay time for you according to the setting of the measurement
if you do not specify a delay. A list of the default for each measurement
function is shown on
Table 4-4
. The range for the delay is from 0 to
3600 seconds. User's selection of delay time is stored in a volatile
memory and the default will be restored when the meter is power-off.
The internal settings of automatic trigger delay
The delay time for automatic trigger delay is determined by the selection
of measurement function, range, and the setting of integration time and
AC filter speed.
The following table shows the default delay time
for each measurement setting.
Table 4-4
Measurement Function
Setting
Trigger Delay
Time
DCV/DCI
PLC >= 1
1.5 ms
PLC < 1
1.0 ms
Ω2 and Ω4
(PLC >= 1)
100Ω ~ 100kΩ
1.5 ms
1 MΩ
15 ms
10 MΩ ~ 100 MΩ
100 ms
Ω2 and Ω4
(PLC < 1)
100Ω ~ 100kΩ
1.0 ms
1 MΩ
10 ms
10 MΩ ~ 100 MΩ
100 ms
ACV/ACI
(Remote Interface/
External Trigger/
Single Trigger)
3 Hz
7.0 s
20 Hz
1.0 s
200 Hz
600 ms
ACV/ACI
3 Hz
1.5 s