138
138
138
138
Frequency and Period Characteristics
Accuracy ± (% of reading)
5
Function
Range
6
Frequency (Hz)
1 Year
23º C±5º C
Frequency
&
Period
100mV
to
750V
3-5
0.10
5-10
0.05
10-40
0.03
40-300K
0.01
AC Characteristics
Accuracy ± (% of r % of range)
7
Function
Range
8
Resolution
Frequency
(Hz)
1 Year
(23°C ± 5°C)
ACV (AC
TRMS
Voltage)
9
100.0000mV
0.1 uV
3-5
1.00 + 0.04
5-10
0.35 + 0.04
10-20K
0.06 + 0.04
5
The Specifications are under the conditions, 6 ½ digits and 2-hour
warm-up.
6
Measurement over 20% at all ranges is allowable except the range
750 VAC.
7
The Specifications are under the test conditions, 6 ½ digits and 2-hour
warm-up, slow AC filter (3 Hz Bandwidth), sine wave input.
8
Measurement over 20% at all ranges is allowable except the range
750 VAC.
9
The Specifications are under the test condition, sine wave input>5%
of range. When the inputs are within 1% to 5% of range and<50 KHz,
the error, 0.1% of range must be added. For inputs within 50 KHz to
100 KHz, the 0.13% of range must be added.