![Picotest M3500A Скачать руководство пользователя страница 90](http://html1.mh-extra.com/html/picotest/m3500a/m3500a_user-manual_1555483090.webp)
90
90
90
90
Remote Interface Operation
Use the following commands to enable and configure MX+B function:
CALCulate:FUNCtion MXB
CALCulate:STATe {OFF|ON}
CALCulate:STATe?
CALCulate:MXB:MMFactor {<
value
>|MINimum|MAXimum}
CALCulate:MXB:MMFactor? [MINimum|MAXimum]
CALCulate:MXB:MBFactor {<
value
>|MINimum|MAXimum}
CALCulate:MXB:MBFactor? [MINimum|MAXimum]
4.3.7 dB/dBm
A. dB
Definition
The dB feature takes a DC or AC voltage measurement and displays it in
decibel unit in correspondence to a relative reference value. The
calculation of dB is listed below:
or
dB = (Input signal in dBm) – (relative value in dBm)
※
NOTE: The
in
V
is
the input signal and the
ref
V
is the relative reference.
The dB measurement is applied to DC and AC voltage only. The relative
value is adjustable and is stored in a volatile memory. The range for the
relative value is between 0
dBm and
200 dBm
.
How to set the relative value
You can either manually select a number for the relative reference or let
the multimeter
take the first reading
(0 dB).
Front Panel Operation
Press SHIFT+ NULL buttons, and then use
◁
and
▷
to move
through the digits and
△
and
▽
buttons to increase or decrease the
numbers to your desired value. Press ENTER to set the value. The
)
log(
20
Vref
V
dB
in
×
=
Содержание M3500A
Страница 1: ...PICOTEST M3500A 6 5 Digit Digital Multimeter User s Manual Version 1 06 ...
Страница 2: ...2 2 2 2 M3500A DMM User s Manual ...
Страница 49: ...49 49 49 49 Figure 3 17 1 2 3 5 5 4 4 1 2 3 ...
Страница 50: ...50 50 50 50 Figure 3 18 5 ...
Страница 191: ...191 191 191 191 FreeLibrary hUSBTMCLIB hUSBTMCLIB NULL return ...