errors found, the error annunciator will be lit to inform you. The
multimeter stores errors using the first-in-first-out (FIFO) order and it
records up to 20 errors in the error queue. The first error detected will be
the first error retrieved. (Refer to Chapter 6 for the list of error
messages.)
1.
The error annunciator will be turned off when you have read the
errors in the memory.
2.
If there is no error found when you check the error queue, the
multimeter shows “
NO ERRORS
”.
3.
If there are more than 20 errors found, the last error in the error
queue will be replaced with “-350”, means too many errors.
4.
The error queue is cleared when the multimeter is turned off, or when
a *CLS command is received.
How to check the error queue
Press MENU then use
◁
and
▷
to locate “SYSTEM” submenu. Press
ENTER to select it. Again use
◁
and
▷
to locate “ERROR” option. Press
ENTER to select it. The message about error queue will show on the
display. Refer to Chapter 6 for the descriptions of all the error
messages. The locations of the buttons are shown with red rectangle
frames in
Figure 4-35
.
Figure 4-35
4.4.10 Firmware Revision
M3500A has three microprocessors for various internal systems. You can
query the multimeter to determine which revision of firmware is installed
Содержание M3500A
Страница 1: ...PICOTEST M3500A 6 5 Digit Digital Multimeter User s Manual Version 1 06 ...
Страница 2: ...2 2 2 2 M3500A DMM User s Manual ...
Страница 49: ...49 49 49 49 Figure 3 17 1 2 3 5 5 4 4 1 2 3 ...
Страница 50: ...50 50 50 50 Figure 3 18 5 ...
Страница 191: ...191 191 191 191 FreeLibrary hUSBTMCLIB hUSBTMCLIB NULL return ...