Component
Description
Transducer handle
Deflection controls
X7-2t Deflection Controls
The smaller knob controls medial/lateral movement, while the larger knob controls anterior/
posterior movement. To place the tip of the TEE transducer into the neutral position, align the
ribs on each knob with the center of the array rotation buttons (as shown below).
The knobs can be controlled by a detent brake that holds the tip position without locking it in
place. This allows the tip to straighten if it meets additional resistance. When the detent brake
actuator is rotated to the right (as shown) both knobs are in the freewheeling mode. When the
detent brake actuator is centered, the small knob (medial/lateral movement) is in the detent
mode, and when the actuator is rotated to the left, both knobs are in the detent mode.
Using the X7-2t TEE Transducer
Transesophageal Transducers
EPIQ 7 User Manual 4535 617 25341
211
Содержание epiq 7
Страница 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Страница 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Страница 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Страница 100: ...Brake Steering Lock Pedal System Overview System Components 100 EPIQ 7 User Manual 4535 617 25341 ...
Страница 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Страница 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Страница 198: ...Intraoperative Transducers Leakage Current Testing for Intraoperative Transducers 198 EPIQ 7 User Manual 4535 617 25341 ...
Страница 244: ...Endocavity Transducers Biopsy with Endocavity Transducers 244 EPIQ 7 User Manual 4535 617 25341 ...
Страница 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...