CAUTION
Make contact with a metal surface of the system before connecting a transducer to the
system.
CAUTION
The following precautions can help to reduce ESD: anti-static spray on carpets; anti-static
spray on linoleum; anti-static mats; or a ground wire connection between the system and the
patient table or bed.
CAUTION
On connectors labeled with the ESD sensitivity symbol
, do not touch the connector pins,
and always observe the preceding ESD precautions when handling or connecting transducers.
NOTE
Electrostatic discharges (ESDs) may cause the ECG heart rate display to increase by 10% to 15%
for a few seconds after the discharge. However, the ECG heart rate display will return to normal
within 4 seconds.
Electromagnetic Emissions
The system is intended for use in the electromagnetic environment specified in the table. The
customer or the user of the system should ensure that it is used in such an environment.
Electromagnetic Compatibility
Safety
EPIQ 7 User Manual 4535 617 25341
71
Содержание epiq 7
Страница 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Страница 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Страница 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Страница 100: ...Brake Steering Lock Pedal System Overview System Components 100 EPIQ 7 User Manual 4535 617 25341 ...
Страница 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Страница 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Страница 198: ...Intraoperative Transducers Leakage Current Testing for Intraoperative Transducers 198 EPIQ 7 User Manual 4535 617 25341 ...
Страница 244: ...Endocavity Transducers Biopsy with Endocavity Transducers 244 EPIQ 7 User Manual 4535 617 25341 ...
Страница 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...