NOTE
If you use cleaning wipes, it may be unnecessary to rinse the transducer with water. Always
follow your product label recommendations.
3. Examine the device and cable for damage such as cracks, splitting, sharp edges, or
projections. If damage is evident, discontinue use of the device and contact your Philips
representative.
Transducer Disinfection by Immersion
Methods for disinfecting transesophageal (TEE) and non-TEE transducers differ. Observe the
following warnings and cautions and then see the procedure for your transducer. For non-TEE
transducers, see
“Disinfecting Non-TEE Transducers by Immersion” on page 268
“Disinfecting TEE Transducers by Immersion” on page 270
Transducers with an AER” on page 273
.
WARNING
If a premixed disinfectant is used, be sure to observe the expiration date.
NOTE
All imaging transducers are rated at a minimum IPX7 in accordance with IEC529. This rating
indicates that the device is protected against the effects of immersion. Continuous wave
(nonimaging) transducers are rated IPX1. This rating indicates that the device is protected
against the effects of vertically falling water. For all TEE transducers, the control module is rated
IPX1, while the endoscope is rated IPX7, in accordance with IEC529.
CAUTION
Using non-recommended disinfectants, using incorrect solution strengths, or immersing a
transducer deeper or longer than recommended can damage or discolor the transducer and
voids the transducer warranty.
Transducer Care
Transducer Disinfection
266
EPIQ 7 User Manual 4535 617 25341
Содержание epiq 7
Страница 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Страница 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Страница 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Страница 100: ...Brake Steering Lock Pedal System Overview System Components 100 EPIQ 7 User Manual 4535 617 25341 ...
Страница 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Страница 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Страница 198: ...Intraoperative Transducers Leakage Current Testing for Intraoperative Transducers 198 EPIQ 7 User Manual 4535 617 25341 ...
Страница 244: ...Endocavity Transducers Biopsy with Endocavity Transducers 244 EPIQ 7 User Manual 4535 617 25341 ...
Страница 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...