NOTE
U
T
is the AC power voltage before application of the test level.
NOTE
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Electromagnetic Immunity: Environment Guidance
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment Guidance
Electrostatic discharge
(ESD), IEC 61000-4-2
± 6 kV contact, ± 8 kV air
Same as IEC 60601 test
level
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast transient/
burst, IEC 61000-4-4
± 2 kV for power supply
lines, ± 1 kV for input/
output lines
Same as IEC 60601 test
level
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge, IEC 61000-4-5
± 1 kV differential mode,
±2 kV common mode
Same as IEC 60601 test
level
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions, and voltage
variations on power supply
input lines, IEC 61000-4-11
Drop out <40 msec
Same as IEC 60601 test
level
Mains power quality should
be that of a typical
commercial or hospital
environment. If you require
continued operation during
power mains interruptions,
Philips recommends that
the system be powered
from an uninterruptible
power supply or a battery.
Electromagnetic Compatibility
Safety
EPIQ 7 User Manual 4535 617 25341
75
Содержание epiq 7
Страница 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Страница 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Страница 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Страница 100: ...Brake Steering Lock Pedal System Overview System Components 100 EPIQ 7 User Manual 4535 617 25341 ...
Страница 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Страница 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Страница 198: ...Intraoperative Transducers Leakage Current Testing for Intraoperative Transducers 198 EPIQ 7 User Manual 4535 617 25341 ...
Страница 244: ...Endocavity Transducers Biopsy with Endocavity Transducers 244 EPIQ 7 User Manual 4535 617 25341 ...
Страница 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...