Inspect the transducer cable, case, and lens before each use. Check for cracks or other damage
that jeopardizes the integrity of the transducer. Report any damage to your Philips customer
service representative, and discontinue use of the transducer.
For all information on transducer cleaning and disinfection, see the
section.
For all information about the use of acoustic coupling gels, see
“Ultrasound Transmission Gels”
.
CAUTION
Some ultrasound coupling gels, as well as some solutions for precleaning, disinfecting, and
sterilizing can damage a transducer. Before using a gel or solution on a transducer, see
“Ultrasound Transmission Gels” on page 184
contact your local Philips representative. For contact information, see
.
If you encounter poor image quality or transducer problems, see
Printer and DVD Recorder Maintenance
Before performing any maintenance on a device, observe the following warnings and cautions:
WARNING
If the device is internal to the system, turn off the system and disconnect the system from the
wall outlet. If the device is external to the system, disconnect the device from the wall outlet.
CAUTION
Do not scratch the roller or allow dirt and dust to contact the roller of a printer.
CAUTION
Do not use strong solvents such as thinner or acetone, or abrasive cleaners, because those
will damage a device cabinet.
System Maintenance
Printer and DVD Recorder Maintenance
294
EPIQ 7 User Manual 4535 617 25341
Содержание epiq 7
Страница 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Страница 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Страница 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Страница 100: ...Brake Steering Lock Pedal System Overview System Components 100 EPIQ 7 User Manual 4535 617 25341 ...
Страница 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Страница 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Страница 198: ...Intraoperative Transducers Leakage Current Testing for Intraoperative Transducers 198 EPIQ 7 User Manual 4535 617 25341 ...
Страница 244: ...Endocavity Transducers Biopsy with Endocavity Transducers 244 EPIQ 7 User Manual 4535 617 25341 ...
Страница 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...