Troubleshooting the Top Cover
7 Troubleshooting
77
If after opening the top cover, or following a repair, the display, all lamps, and all function buttons fail to
work, check the ribbon cable connection to the MIF board.
The top cover contains the following items:
• Charge/communication contacts (see “Checking Contacts” on page 65)
• Function buttons
• Display/window
• LEDs
• Magnets
Function Buttons
If any of the buttons fail to function, reset the system (see “System Reset” on page 61). If an individual
button fails to work, try to solve the problem by renewing the top cover (see “Removing the Top Cover
Assembly” on page 20 and “Replacing the Top Cover Assembly” on page 21). If all the function buttons
fail to work, then this is likely to be a MIF board malfunction, a repair being effected by exchanging the
base station.
Содержание Avalon CTS M2720A
Страница 2: ...ii ...
Страница 22: ...1 General Information Antenna System Guidelines 16 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 28: ...2 Theory of Operation ECG Frontend Hardware 22 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 32: ...3 System Interfaces Compatible Fetal Monitors 26 ...
Страница 52: ...5 Spare Parts Antenna and Base Station Part Numbers 46 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 66: ...6 Preventive Maintenance and Safety Battery Exchange 60 ...
Страница 92: ...9 Upgrades 86 ...
Страница 102: ...10 Specifications Essential Performance 96 ...
Страница 104: ...A Removing and Replacing the Transducer Battery 98 Wipe off excess silicone grease ...
Страница 106: ...B Removing and Replacing the Transducer Belt Button 100 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 108: ...C Avalon CTS Frequency Table 102 ...