Essential Performance
10 Specifications
95
The second table identifies Essential Performance in the presence of “transient” interference:
• Electrostatic Discharge (ESD)
• Electrical Fast Transients/Bursts
• Surges
• Voltage interruptions
• Electro Surgery (ESU)
Parameter
Essential Performance
General
No interruption or cessation of current operating mode (e.g., no reboot, display is functional,
etc.)
No spontaneous operation of controls (e.g., no activation of touch screen controls without
user interaction)
ECG
Measurement of the heart rate within the specified error limits and over the specified
measurement ranges
Alarming on heart rate limit violation within specified delay time.
Toco Extern
Measurement of the external Toco within specified error limits and over the specified
measurement range
Ultrasound
Measurement of the fetal heart rate within the specified error limits and over the specified
measurement ranges
Alarming on heart rate limit violation within specified delay time.
Parameter
Essential Performance
All
After electrostatic discharge, fast transients/bursts, surges and electro surgery interference, the
equipment will return to previous operation mode within 10 seconds without loss of any stored
data.
After voltage interruptions the equipment returns to previous state without operator intervention
and loss of data.
Содержание Avalon CTS M2720A
Страница 2: ...ii ...
Страница 22: ...1 General Information Antenna System Guidelines 16 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 28: ...2 Theory of Operation ECG Frontend Hardware 22 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 32: ...3 System Interfaces Compatible Fetal Monitors 26 ...
Страница 52: ...5 Spare Parts Antenna and Base Station Part Numbers 46 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 66: ...6 Preventive Maintenance and Safety Battery Exchange 60 ...
Страница 92: ...9 Upgrades 86 ...
Страница 102: ...10 Specifications Essential Performance 96 ...
Страница 104: ...A Removing and Replacing the Transducer Battery 98 Wipe off excess silicone grease ...
Страница 106: ...B Removing and Replacing the Transducer Belt Button 100 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 108: ...C Avalon CTS Frequency Table 102 ...