6 Preventive Maintenance and Safety
Regular Preventive Maintenance
56
Visual Check
Ensure there are no cracks in the transducer housing. Make sure the battery drawer fits properly into the
transducer housing, and that the sealing lip is in good condition. Inspect the LCD window, and if there is
water or condensation behind the window, replace the transducer. On the ECG transducer, ensure that
the adapter cable socket is not damaged.
Toco Ventilation/Belt Button
The Toco transducer requires air exchange for drift-free operation (see also the section “Underwater
Monitoring” in the
Instructions for Use)
. This is done by a ventilation membrane incorporated in the belt
button. The effectiveness of the ventilation membrane should be checked at least once a year to ensure
that free air exchange takes place. If the membrane is congested, the belt button must be exchanged (see
the instruction sheet, “Removing and Replacing the Transducer Belt Button” that accompanies the belt
button replacement kit, or see page 91 in this manual).
Testing the Ventilation Membrane
If you want to test the ventilation membrane in the belt button, you need the following:
• Manometer.
• Test volume chamber.
• Three lengths of silicone tubing with a ‘T’ adapter.
1
Remove the belt button from the transducer using the special tool (see the instruction sheet “Removing
and Replacing the Transducer Battery/Belt Button”).
2
Connect the manometer to the volume chamber using two lengths of the silicone tubing and ‘T’
adapter as shown. Connect one end of the remaining length of silicone tubing to the ‘T’ adapter, and
the other to the threaded end of the belt button.
3
Apply pressure with the manometer (for example, 250 to 260 mmHg).
4
Release the manometer valve. The pressure should fall rapidly, to indicate that the membrane is
functioning correctly.
‘T’ adapter
Belt button
Manometer
Silicone
tubing
Содержание Avalon CTS M2720A
Страница 2: ...ii ...
Страница 22: ...1 General Information Antenna System Guidelines 16 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 28: ...2 Theory of Operation ECG Frontend Hardware 22 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 32: ...3 System Interfaces Compatible Fetal Monitors 26 ...
Страница 52: ...5 Spare Parts Antenna and Base Station Part Numbers 46 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 66: ...6 Preventive Maintenance and Safety Battery Exchange 60 ...
Страница 92: ...9 Upgrades 86 ...
Страница 102: ...10 Specifications Essential Performance 96 ...
Страница 104: ...A Removing and Replacing the Transducer Battery 98 Wipe off excess silicone grease ...
Страница 106: ...B Removing and Replacing the Transducer Belt Button 100 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 108: ...C Avalon CTS Frequency Table 102 ...