79
8
8
Changing Settings
There are four classes of settings for individually configuring the Avalon CTS:
• Bed Label setting (refer to
Instructions for Use
)
• C = Configuration settings (refer to
Instructions for Use
)
• F = Function settings
• A = Action settings
The C and F settings can be set:
– using a PC with the Service Support Tool installed. This is the recommended method.
– manually, using the keys on the base station.
The A settings can only be set manually, using the keys on the base station.
Configuration Settings
The
Instructions for Use
describes how to manually change the theft protection level (C1), theft protection
alert volume (C2), key click volume (C3), audible alert volume (C4) and acoustical alarmdefault (C5).
To Purchase, Visit
Содержание Avalon CTS M2720A
Страница 2: ...ii ...
Страница 22: ...1 General Information Antenna System Guidelines 16 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 28: ...2 Theory of Operation ECG Frontend Hardware 22 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 32: ...3 System Interfaces Compatible Fetal Monitors 26 ...
Страница 52: ...5 Spare Parts Antenna and Base Station Part Numbers 46 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 66: ...6 Preventive Maintenance and Safety Battery Exchange 60 ...
Страница 92: ...9 Upgrades 86 ...
Страница 102: ...10 Specifications Essential Performance 96 ...
Страница 104: ...A Removing and Replacing the Transducer Battery 98 Wipe off excess silicone grease ...
Страница 106: ...B Removing and Replacing the Transducer Belt Button 100 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 108: ...C Avalon CTS Frequency Table 102 ...