Specifications
Specifications for the accuracy of heart rate measurements are given in the
Instructions for Use
of the fetal
monitor.
General
Base Station
Environmental Specifications (Transducers and Base Station)
Temperature Range
Charging
0°C to 45°C (32°F to 113°F)
Operating
0°C to 45°C (32°F to 113°F)
Storage (without battery)
-20°C to +60°C (-4°F to 158°F)
Storage with battery
Depends on initial charge level and temperature
(storage time decreases significantly at high (> 45°C/
113°F) temperatures
Humidity Range
Operating
5% to 95% relative humidity @ 40°C/104°F
Storage
5% to 85% relative humidity @ 50°C/122°F
Altitude Range
Operating
3000m/9800 ft.
Storage
15000m/49000 ft.
Base Station Specifications
Receiver Unit
Power
Supply Voltages
100 VAC to 240 VAC ± 10%
Supply Frequency Range
50 Hz to 60 Hz
Consumption
15 VA
Type of Protection Against
Electrical Shock
Class I equipment
Dimensions and Weight
Size mm/(in): width x depth x height
350 x 240 x 75 (13.8 x 9.5 x 3.0 in)
Weight
2.5kg/5.5 lbs without transducers
Input Sensitivity
Input Sensitivity
-110 dBm @ 30dB Signal-to-Noise Ratio
Image Rejection
Image Rejection
> 80dB
Ranges
Frequency Range
See Frequency options
Receiving Range (line of sight)
approximately 100m/300 ft.
Содержание Avalon CTS M2720A
Страница 2: ...ii ...
Страница 22: ...1 General Information Antenna System Guidelines 16 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 28: ...2 Theory of Operation ECG Frontend Hardware 22 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 32: ...3 System Interfaces Compatible Fetal Monitors 26 ...
Страница 52: ...5 Spare Parts Antenna and Base Station Part Numbers 46 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 66: ...6 Preventive Maintenance and Safety Battery Exchange 60 ...
Страница 92: ...9 Upgrades 86 ...
Страница 102: ...10 Specifications Essential Performance 96 ...
Страница 104: ...A Removing and Replacing the Transducer Battery 98 Wipe off excess silicone grease ...
Страница 106: ...B Removing and Replacing the Transducer Belt Button 100 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 108: ...C Avalon CTS Frequency Table 102 ...