4 Disassembly/Reassembly
Changing the Power Supply/Metal Chassis Assembly
38
Step 6: Replacing the MIF Board
1
Carefully locate the multipin connector plug, on the underside of the MIF board, in the socket on the
receiver board, ensuring it is accurately aligned.
2
Carefully apply a little pressure on the left hand rear corner of the MIF board to seat the multipin
connector into the socket on the receiver board below.
3
Ensure that the screw holes in the board align with the screw fixing points below.
4
Drive in the two screws (
B
) with a T-10 Torx driver.
5
Then drive in the two screws (
A
) using a T-10 Torx driver. The MIF board is now secure.
Multipin connector
Screws (
B
)
Screws (
A
)