Disassembly/Reassembly
WARNING
• Performance verification: do not place the system into operation after repair or maintenance has been
performed, until all performance tests and safety tests listed in Chapter 6 of this service manual have been
performed. Failure to perform all tests could result in erroneous parameter readings, or patient/operator
injury.
• Energized circuits are accessible with the covers open. Do not work on the base station with the covers open
and AC power connected. Only qualified service personnel should open or disassemble the base station.
• Before attempting to open or disassemble the base station, disconnect it from the AC mains supply.
CAUTION
Observe ESD (electrostatic discharge) precautions when working within the unit.
Introduction
Remember to store all screws and parts in a safe place for later refitting.
How to Use this Chapter
The disassembly sections detail the step-by-step procedures you use to access replaceable parts of the base
station.
All the transducers are sealed at the factory and are not repairable. Replacement transducers are supplied
complete with both battery and belt button fitted.
The Avalon base station consists of two major assemblies:
• The top cover assembly
• The chassis assembly
The top cover assembly contains the human interface (HIF) board, the keypad inlay, and the battery charge
contacts and wiring. The top cover assembly is only available as a complete assembly, and a repair to any part
of the top cover assembly is effected by replacement of the whole assembly.
The chassis assembly contains the metal chassis and power supply, the receiver board and the monitor
interface (MIF) board. The power supply always comes complete with the chassis (without boards), and a
repair to either the power supply or the chassis is effected by replacement of this chassis/power supply
assembly.
Содержание Avalon CTS M2720A
Страница 2: ...ii ...
Страница 22: ...1 General Information Antenna System Guidelines 16 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 28: ...2 Theory of Operation ECG Frontend Hardware 22 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 32: ...3 System Interfaces Compatible Fetal Monitors 26 ...
Страница 52: ...5 Spare Parts Antenna and Base Station Part Numbers 46 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 66: ...6 Preventive Maintenance and Safety Battery Exchange 60 ...
Страница 92: ...9 Upgrades 86 ...
Страница 102: ...10 Specifications Essential Performance 96 ...
Страница 104: ...A Removing and Replacing the Transducer Battery 98 Wipe off excess silicone grease ...
Страница 106: ...B Removing and Replacing the Transducer Belt Button 100 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 108: ...C Avalon CTS Frequency Table 102 ...