  Interval Time (min)= sputter increments, in 

units of .5min. 

  Delayed Acq. Time (sec) = Delay before 

acquisition to allow count rate to stabilize. 3s is 
typical. 

  Num(Presputter) = How many scans to take 

before 1

st

 sputter. 1 is typical. 

o

  Now, number of cycles is determined by sputter settings. Add 

“repeats” to increase signal for some acquisitions. 

o

  Push 

ACQUIRE

 to begin depth profile. 

o

  The charge neutralizer can be used during depth profiling. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

5.  To Shut Down Sputtering/Neutralizer 

a. 

Summitt->

 Acquisition Setting window (LHS)-> Choose 

No Sputter

 

in “Sputter Mode” category 

(FIG 9)

b. 

Summitt->

 Acquisition Setting window (LHS)-> Choose 

Off

 in 

“Neutralize” category 

(FIG 9)

c. 

Summitt-> 

“Image” Window (RHS) -> “Ion Gun” Tab: Push 

Off (FIG 

6)

 

d.  On Pfeiffer RVG050 box 

(FIG 7)

, toggle white switch to 

Limit

.  

Never turn off power to Pfeiffer RVG050! 

 

e. 

Summitt->

 “Image” Window -> Push 

Off 

in Neutralizer Tab 

(FIG 8)

FIG 9. Depth Profile Setup 

Содержание VersaProbe

Страница 1: ...tart contact a lab supervisor Make sure Summitt is not already open and minimized before opening Two open versions will cause problems Never close Watcher It remains open even if Summitt is restarted...

Страница 2: ...a new timer starts for the turbo Wait for timer to finish and Task Pump Intro complete message in Watcher The gauge for the Intro chamber will read until 10 2 Pa and then read 0 While waiting for intr...

Страница 3: ...3 When stage has stopped moving follow instructions in Intro dialog box see next step below FIG 1 Vacuum Watcher Intro Procedure FIG 2 Summit Image Window Stage Tab Intro Procedure 3 5 8...

Страница 4: ...older The Intro dialog box will indicate that the stage is moving 6 If the stage has successfully grabbed the sample and is no longer moving carefully pull the transfer rod out of the chamber to the s...

Страница 5: ...maximum The stage height will rise for routine increments and then decrease and rise again in smaller steps until done Push Ok and close AutoZ window when finished a If final height is not on desired...

Страница 6: ...est statistics and most counts but may need smaller spot for localized area or acquisition within sputtered region See Figure 5 6 Summitt Acquisition Settings Window LHS Push Acquire to start 7 To use...

Страница 7: ...ing you to push Diff vlv Open OK e On Pfeiffer RVG050 FIG 7 box near computer monitor toggle white switch to Set Point In Ion Gun tab check Ext Pressure mPa readout and watch until stabilizes 10mPa Th...

Страница 8: ...utter Modes Manual Sputter o Summitt When pressure conditions are ready pushing Sputter in Ion Gun tab will immediately sputter o For manual timed sputter o choose a low KV sputter file check timed bo...

Страница 9: ...ACQUIRE to begin depth profile o The charge neutralizer can be used during depth profiling 5 To Shut Down Sputtering Neutralizer a Summitt Acquisition Setting window LHS Choose No Sputter in Sputter M...

Страница 10: ...indow will open prompting to insert sample Carefully push transfer rod into chamber and sliding it until the stop position Fork should slide onto sample holder Push OK on Task Message Window 3 Summitt...

Страница 11: ...11 FIG 11 Vacuum Watcher Extract procedure 2 4 5...

Отзывы: