7
b. For sputtering, choose desired sputter file. Chart taped to PHI
console explains choices.
c. Push
Standby
(will turn on power to sputter gun).
d. A dialogue box will open reminding you to push Diff vlv Open.
OK.
e. On Pfeiffer RVG050
(FIG 7)
box near computer monitor, toggle
white switch to
Set Point
. In “Ion Gun” tab, check
Ext.Pressure(mPa) readout and watch until stabilizes ~10mPa.
The main chamber pressure will also get worse (1x10
-6
).
FIG 6: Summit “Image” Window, Ion Gun
FIG 7: Pfeiffer RVG050C Vacuum Control Box
Choose sputter settings
Choose Neutralize file