4
4.
Watcher ->
Push
Transfer Sample
•
A “
Task Message
” window will open. Press
Yes
to confirm sample
transfer.
•
Gate valve (V1) will open. Wait until white line has moved to “open”
position on gate valve to confirm before moving sample.
•
Another “
Task Message
” window will open prompting to insert
sample. Carefully push
transfer rod
into chamber by grasping
black handle and sliding it until the stop position. Push
OK
on Task
Message Window. Sample should be on fork, positioned just
above the stage in the main chamber.
5.
Summitt ->
Push
Up [Z]
in the
“Intro” dialog box
. The stage will move Z
upward to grab the sample holder. The
“Intro” dialog box
will indicate that
the stage is moving.
6. If the stage has successfully grabbed the sample and is no longer moving,
carefully pull the
transfer rod
out of the chamber to the stop position.
7. The gate valve (V1) will automatically close.
If stage is not aligned to the sample correctly, and a problem arises
during Up [Z], push
Abort
in the
“Intro” dialog box
and manually
drop Z. To drop Z:
Summitt “Image” Window->
type a small
number in the Z(mm) box on the “Stage” tab and push
Move
.
Contact a lab manager for help.
8.
Summitt ->
In the “Intro” dialog box, push
Set
(Sample Inf.) to update
sample thickness and holder size, thereby enabling collision protection for
the extractor lens and good Z-height alignment. Push
OK
to close box.
FIG 3: Summit “Image” Window, Auto Z FIG 4: Stage Controller