12
INSTALLATION
1.
After unpackaging
Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the
outside panels are dirty, clean them with a diluted neutral dishwashing detergent.
Undiluted detergent can damage the plastic components. For the dilution, refer to the instruction of
the detergent.
After the cleaning with the diluted detergent, always wipe it off with a wet cloth. Then wipe off the
panels with a dry cloth.
Note:
Remove the cable tie banding the power supply cord. Prolonged banding may cause the corrosion of
the cord coating.
2.
Installing the unit
Extend the leveling feet by rotating them to separate 2
front casters from the floor. [Fig. 1]
And adjust the leveling foot to level the unit.
3. Fixing the unit
Fix the unit to the wall by using 2 back spacers [Fig. 2] and
a rope or chain.
The back spacers (which are also used to prevent the unit from
toppling over) have a dual purpose: To provide a clearance
between the unit and the wall behind it, and to prevent the unit
toppling over.
4. Ground (earth) the unit
The ground (earth) is for preventing the electric shock in
the case of the electrical insulation is somehow degraded.
Always ground the unit at the time of installation.
A 3-prong plug with grounding pole is provided to the
poser supply cord. There is no need for electric work for
grounding.
If the power supply outlet is not grounded, it will be
necessary to install a ground by qualified engineers.
Leveling foot
Fig. 1
Fig. 2
Back spacer