34
INSTALLATION OF TEMPERATRUE RECORDER
Attachment of MTR-0621LH
1.
Remove the bottom shelf (bottom right drawer for
MPR-514R) in the chamber.
For MPR-514, remove the doors before taking out the
bottom shelf.
2.
Unscrew the screw fixing the panel cover at the space
for automatic temperature recorder and open the panel
cover. [Fig. 1]
3.
Remove the panel cover from the shaft by pushing the
shaft edge to the right and left side respectively.
4.
Fix the temperature recorder to the frame in the recorder
fixing (MPR-S30) referring to the instruction manual
enclosed with the recorder fixing. [Fig. 2]
5.
Unscrew 2 screws on the air intake vent cover (right
side) on the front bottom and then remove the air intake
vent cover. [Fig. 3]
<Important>
A sensor is fixed to the back side of the air intake vent
cover. Do not pull the air intake vent cover excessively.
6.
Take out the insulation in the sensor port and pass the
recorder sensor into the chamber through the sensor port
from the space for temperature recorder. [Fig. 4]
Fig. 2
Shaft
Panel cover
Frame
Fig. 1
Screw
Air intake vent cover
<overhead view of right chamber>
<overhead view of right chamber>
Sensor port
Fig. 3
Fig. 4