39
INSTALLATION OF TEMPERATURE RECORDER
7.
Fix the recorder sensor to the sensor fixture. [Fig. 5]
The center of the recorder sensor should be the center
of the sensor fixture.
8.
Replace the insulation to the sensor port and cover the
sensor port with the putty. Then replace and fix the air
intake vent cover by 2 screws.
9.
Connect the recorder connector to the connector for
recorder power located at upper right inside the space for
temperature recorder. [Fig. 6]
The connector for recorder power has a connector cover.
Remove the connector cover before connection.
10.
Set the temperature recorder with the channel on the
recorder fitting aligned with the shaft and fix the
temperature recorder by a screw at the space for
temperature recorder. [Fig. 7]
Then put a cap on the screw removed in procedure 2 on
page 38.
11.
Replace the right bottom shelf (right bottom drawer for
MPR-1014R) in the chamber.
Fig. 5
Recorder sensor
Sensor fixture
Sensor fixture
Fig. 6
Fig. 7
Connector for recorder power
Space for automatic temp. recorder
Screw
Shaft
Channel