INSTALLATION SITE
To operate this unit properly and to obtain maximum performance, install the unit in a location with the
following conditions:
A location not compliance with the following conditions may cause poor performance, failure or
accident.
A location not subjected to direct sunlight
Do not install the unit under direct sunlight. Installation in a location subjected to direct sunlight may
degrade the performance.
A location with adequate ventilation
Leave at least 10 cm around the unit for ventilation. Poor ventilation will result in a reduction of the
performance and consequently the failure.
A location away from heat generating sources
Avoid installing the unit near heat-emitting appliances such as a heater or a boiler etc. Heat can
decrease the intended performance of the unit.
A location with little temperature change
Install the unit under stable ambient temperature. Installing the unit under unstable ambient temperature
may result in unstable performance.
A location with a sturdy and level floor
Always install the unit on a sturdy and level floor withstanding the total weight of the unit. The uneven
floor or tilted installation may cause failure or injury. Unstable condition may cause vibration or noise.
A location not prone to high humidity
Install the unit in the ambient of 80 % R.H. or less humidity. Installation under high humidity may cause
current leakage or electric shock.
CAUTION
The unit may collect excessive frost on the evaporator if it is installed in high temperature and high
humidity location. This will cause frequent defrosting.
A location without flammable or corrosive gas
Never install the unit in a flammable or corrosive location. This may cause explosion or fire or may result
in the current leakage or electric shock by the corrosion of the electrical components.
CAUTION
Never install the unit in a location where the corrosive material such as sulfur compound is likely
to be generated (near a drainage facilities, etc).
Corrosion of the copper pipe may result in
deterioration of cooling unit and consequently the failure.
A location without the possibility of anything fall
Avoid installing the unit in the location where anything can fall down onto the unit. This may cause the
breakdown or failure of the unit.