REMOTE ALARM TERMINAL
The alarm status is noticed to a remote location when a remote alarm equipment (commercial item) is
connected to the remote alarm terminal. It is recommended to install a remote alarm equipment
(commercial item) when the refrigerator is installed in a desolate location so that an alarm status is
noticed to an operator.
Contact our representative or agent for the installation of a remote alarm equipment (commercial item).
Location of remote alarm terminal: rear lower left
Allowable contact capacity: DC 30 V, 2 A
The alarm status of remote alarm terminal is cancelled by pressing the alarm buzzer stop key (BUZZER)
since the remote alarm is operated in conjunction with alarm buzzer.
However, the alarm status of remote alarm terminal is not cancelled by pressing the alarm buzzer stop
key (BUZZER) when the unit is under power failure alarm (with MPR-48B1 installed).
The remote alarm terminal returns to the alarm status again after a certain period (ringback time) if the
same alarm status continues.
The remote alarm terminal is in alarm status when the power supply cord is unplugged because it is
regarded as a power failure. In this case, the alarm status of remote alarm terminal is not cancelled by
pressing the alarm buzzer stop key (BUZZER).
Connection of remote alarm terminal
1.
Remove the terminal cover at rear lower left of the cabinet by unscrewing 2 screws. [Fig. 1]
2.
Connect the lead wire of a remote alarm equipment (commercial item) to the terminal.
It is recommended to use cables with a maximum length of 30 meters.
3.
Replace the terminal cover and fix it by 2 screws.
REMOTE
ALARM
MAX DC30V 2A
N.C. COM N.O.
Terminal
cover
<
Rear lower left
>
Fig. 1