18
ALARM TEMPERATURE SETTING
Setting of high temperature alarm
By setting the high temperature alarm, the alarm indicator and the temperature display blink and the
alarm buzzer operates (after 15 minutes) when the chamber temperature is over the setting of high
temperature alarm. Set the high temperature alarm to protect the stored items against the damage
resulting from temperature rise.
Setting range of high temperature alarm:
Between chamber temperature plus 2
o
C and chamber temperature plus 14
o
C
Initial setting (factory setting): chamber temperature plus 5
o
C
Example: Change the high temperature alarm to chamber temperature plus 3
o
C from chamber
temperature plus 5
o
C
►
Following shows a sample setting. Set the desired high temperature alarm according to the condition
of use.
Description of operation
Key operated
Indication after operation
1
-----
The current chamber temperature is
displayed.
2 Press up arrow key for 5 seconds.
F00 is displayed and the first digit
blinks.
3 Press up arrow key once.
The display is changed to F01 from
F00.
4 Press set key.
SET
The current setting (005) is displayed
and the first digit blinks.
5
Press up arrow key and scroll the
figure to 3.
The display is changed to 003 from
005.
6 Press set key.
SET
Set temperature is memorized and
the current chamber temperature is
displayed.
The setting mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation (auto-return function). In this case, the setting is not accepted.
<Important>
The high temperature alarm may be activated after defrosting or a large amount of load is stored in the
chamber. This is not a malfunction. The high temperature alarm is cancelled automatically when the
chamber temperature reaches the set temperature.