![Phcbi MCO-5M Series Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/phcbi/mco-5m-series/mco-5m-series_operating-instructions-manual_1536748054.webp)
52
SPECIFICATIONS
Product name
O
2
/CO
2
Incubator MCO-5M
Accessories
1 power supply removable cord*
1
, 3 trays, 3 gas tube, 1 humidifying pan
1 gas injection nozzle, 1 gas injection nozzle tube
1 pair of stacking plates A and B, 4 protective stickers, 6 tube bands
Weight 50
kg
Optional accessories
UV system set (MCO-19UVS)
Gas auto changer (MCO-5GC), Gas regulator (MCO-010R)
Tray (MCO-30ST), Roller base (MCO-5RB)
Interface board (MTR-L03)*
2
, Interface board (MTR-480)*
2
Interface board (MCO-420MA, USA only)
Note
: Refer to the updated catalog when ordering an optional component.
Designs and specifications are subject to change without notice.
*1: For the MCO-5M for EU countries, 2 power supply removable cords are supplied with it.
a) Power supply removable cord for UK
b) Power supply removable cord for EU countries other than UK
Refer to page 27.
*2: Only for the Data acquisition system MTR-5000 user.
PERFORMANCE
Product name
O
2
/CO
2
Incubator MCO-5M
Model number
MCO-5M-PT MCO-5M-PA MCO-5M-PK
MCO-5M-PE
Temperature control range
Ambient tempe5
o
C to 50
o
C (ambient temperature; 5
o
C to 35
o
C)
Temperature distribution
±0.25
o
C (ambient temperature; 25
o
C, setting; 37
o
C, CO
2
: 5 %, O
2
: 5 %, no load)
Temperature variation
±0.1
o
C (ambient temperature; 25
o
C, setting; 37
o
C, CO
2
: 5 %, O
2
: 5 %, no load)
CO
2
control range
0 % to 20%
CO
2
variation
±0.15 % (ambient temperature; 25
o
C, setting; 37
o
C, CO
2
: 5 %, O
2
: 5 %, no load)
O
2
control range
1 % to 18 %, 22 % to 80%
O
2
variation
±0.2 % (ambient temperature; 25
o
C, setting; 37
o
C, CO
2
: 5 %, O
2
: 5 %, no load)
Chamber humidity
95 %R.H.± 5 %R.H.
Applicable environment
condition
Temperature; 5
o
C to 35
o
C, Humidity; equal or less than 80 %R.H.
(The designed performance may not be obtained when
the ambient temperature is equal or less than 15
o
C)
Noise level
24 dB (A scale)
Power consumption
Max. 205 W
Heat emission
Max. 740 kJ/h
Rated voltage, frequency
AC 110 V-120 V, 60 Hz
AC 220 V, 60 Hz
AC 220 V-240 V
50 Hz
Amperage
Max. 1.8 A
Max. 0.9 A
Max. 0.9 A
Note:
The unit with CE mark complies with EU directives.
Based on our measuring method.
Содержание MCO-5M Series
Страница 2: ......