Remote alarm terminals
The remote alarm terminal is located at the rear right side.
The remote alarm terminal is a contact output.
Normal
: OPEN
Abnormal , blackout : CLOSE
Contact capacity : DC 30 V, 2 A
Note:
• When the power switch is OFF or the power failure condition, the contact output is CLOSE.
• The remote alarm cannot be silenced by pressing the alarm buzzer stop key (BUZZER) since the
remote alarm is not conjunct with the alarm buzzer stop key (BUZZER) key.
Remote alarm terminal
Remote alarm terminal
REMOTE ALARM
MAX DC30V 2A
N. O.