![Phcbi MCO-5M Series Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/phcbi/mco-5m-series/mco-5m-series_operating-instructions-manual_1536748023.webp)
21
INITIAL CLEANING METHOD
Before using the incubator for the first time, clean dirt (tape residue, smear, etc.) from the
chamber and the inner attachments thoroughly. To keep the chamber clean is essential
to get the proper performance out of the incubator. Use the following steps to clean the
incubator properly.
1.
Remove the inner attachments, referring to “REMOVING INNER ATTACHMENTS
”
on page 22.
2.
Clean the removed inner attachments, chamber inside walls and inner door gaskets with a cloth or
sponge soaked in neutral detergent, diluted by 5 % or less. (Undiluted detergent can damage the plastic
components. For the dilution, refer to the instruction of the detergent.) (Fig. 1)
CAUTION
Do not use detergents or antiseptic solutions with acid, alkali, or chlorine. Doing so may cause
discoloration, corrosion, or rusting.
Be careful to keep the diluted detergent or water out of the temperature sensor, the CO
2
gas injection
hole, the gas injection port, the inner sample air access port, the fan motor shaft bearing, and the inner
sample air outlet (Fig.2 ). Also, do not wash the temperature sensor, the CO
2
sensor, the O
2
sensor,
the water level sensor and UV lamp using detergent. Doing so may cause failure.
(
Fig.2
)
3.
Soak a cloth or unwoven cloth in distilled water and wring it tightly, and then wipe off the residual
detergent thoroughly.
4.
Wash the silicon caps (2 pcs) for the access port and the inner circulating fan using the above
mentioned detergent and rinse them with distilled water, and then autoclave them for sterilization (121
o
C,
20 minutes).
5.
Wipe up the inside walls, and the inner attachments like trays thoroughly with a cloth or unwoven cloth
soaked in alcohol for disinfection. Be careful not to leave any residue alcohol.
6.
Reinstall the inner attachments correctly and securely, referring to “INSTALLING INNER
ATTACHMENTS
”
on page 24.
Fig. 1
Fig. 2
Temp. sensor
Water level sensor
Silicon cap
UV lamp
CO
2
sensor
O
2
sensor
Содержание MCO-5M Series