FILLING THE HUMIDIFYING PAN
Use the following procedure to fill the humidifying pan or to replace the water.
1.
Lift the front side of the humidifying pan cover.
(Fig. 1)
2.
Pull the humidifying pan forward. (Fig. 2)
3.
Dispose of the remaining water in the humidifying
pan and clean the humidifying pan with a diluted
detergent. Then rinse it thoroughly with distilled
water and wipe it with alcohol for disinfection.
4.
Wipe all moisture from the bottom of the chamber.
5.
Return the humidifying pan to the chamber and
add sterile distilled water (approx. 1.5 L, preheated
to 37
C). (Fig. 3)
6.
Set the humidifying pan with the inner side flush
against the back, and replace the humidifying pan
cover. Close inner door and outer door, and confirm
that water level alarm lamp (RH PAN) on the control
panel is OFF.
Note:
• Preheat to 37
C the water to be added to the
humidifying pan. Adding low-temperature water will
lower the temperature and humidity in the chamber.
• Replace the water in the humidifying pan by the
above procedure when the water level alarm lamp
blinks or water in the humidifying pan decreases.
CAUTION
When refilling the water in the humidifying pan,
always wipe off any dirt from the water level sensor
with alcohol for disinfection
. While doing that, be careful not to apply excessive force to the sensor lead
wire.
Caution:
In order to adjust the humidity inside the incubator chamber optimally and prevent condensation on the
chamber surface and inner door, it is designed to create the lower temperature area underneath the
humidifying pan to recondense the evaporated moisture in the humidifying pan. Although the
condensation may be seen round the humidifying pan on the bottom of chamber (inside the humidifying
pan cover), it is not a malfunction.
Fig. 1
Fig. 2
Fig. 3