9

4. Operations 

4.1. Before Switching ON

Turbopump rotors rotate at high speed. When

the high vacuum flange is open there is a danger
of injury and of damage to the pump caused by
objects falling into the pump. Therefore never
operate the pump with open high vacuum flange.

When operating with casing heating unit or TMS

operations phigh temperatures can arise on the
pump surfaces.

Pre-select the required operational functions and parame-
ters on the Magnetic Bearing Controller TCM 1601.

With water cooling: Open cooling water supply and check
flow.

4.2. Switching ON

Switch on the magnetic bearing electronic controller with
switch S1 on the rear side of the TCM 1601.

Switch on the pumping station with the push button 

on

the front panel of the TCM 1601.

– If air cooling is involved the cooling fan is also switched on

by the magnetic bearing controller.

– The backing pump is automatically started with the

Pumping Station Control Unit TCS 180.

Switch on the backing pump if it is not switched via the
pumping station control unit.

Take care when pumping hazardous gases and
observe the safety precautions of the gas manu-
facturer.

4.3. Heating (only for pumps with casing 

heating unit)

The heating duration is dependent on the degree of contami-
nation and on the required final pressure. The heating should
last for at least 4 hours.

Where casing heating unit is involved the turbo-

pump must be water cooled.

Switch on the  turbopump heating in the menu of the
Magnetic Bearing Controller TCM 1601 (please see
Operating Instructions PM 800 478 BN for the TCM 1601).

PLEASE NOTE

3.9. Connecting The Temperature 

Management System (TMS)

Pumps with the designation TMH 1600M T/TMU 1600M T
(please see the rating plate) can be heated to a maximum of
90 °C in the Holweck stages range with the heaters and tem-
perature sensors (TMS) integrated in the middle casing.
During TMS operations the additional flange heating for the
turbomolecular pump is not possible.

During TMS operations the  turbopump must be

operated with water cooling.

For connection and operation of the temperature
management system (TMS) please refer to operating
instructions for the TMS, PM 800 517 BN and 
for the TCM 1601, PM 800 478 BN.

PLEASE NOTE

Temperature management system

95 Connection box
96 Connecting cable for the heating elements

96

95

Содержание TMH 1600M

Страница 1: ...Operating Instructions Turbomolecular Drag Pumps With Magnetic Bearings TMH 1600M TMU 1600M PM 800 344 BE C 0011...

Страница 2: ...cting The Casing Heating Unit 8 3 8 Connecting The Magnetic Bearing Controller 8 3 9 Connecting The Temperature Management System TMS 9 4 Operations 9 4 1 Before Switching ON 9 4 2 Switching ON 9 4 3...

Страница 3: ...opump When returning the turbopump observe the shipping instructions Use at least 6 bracket screws to connect the high vacuum flange The turbopump must be anchored in accordance with the installation...

Страница 4: ...g and sensors are accessible cen trally via a separate connection box When the TMS is in operation the additional casing heating unit on the turbomolecular pump is not possible 2 Understanding The TMH...

Страница 5: ...e volume flow rate of the pump by approximately 15 For fitting splinter shields please refer to Fitting the splinter shield The high vacuum side can be flanged directly to the vacuum chamber or via a...

Страница 6: ...g pump Please see the operating instructions for the Magnetic Bearing Controller TCM 1601 PM 800 478 BN 3 4 Connecting The Cooling Unit The Turbopumps TMH TMU 1600M are water cooled as stan dard The p...

Страница 7: ...M turbopump see the operating instructions for the TCM 1601 PM 800 478 BN Alternatively either the casing heating unit or the TMS can be connected to the control line Simultaneous operations of the fu...

Страница 8: ...ted during installation When baking out do not touch the heating sleeve pump casing and vacuum chamber Screw the heating sleeve tightly onto the casing with the securing screws and fit protection agai...

Страница 9: ...ing hazardous gases and observe the safety precautions of the gas manu facturer 4 3 Heating only for pumps with casing heating unit The heating duration is dependent on the degree of contami nation an...

Страница 10: ...somewhat lower frequency 4 4 Operations With The TMS If the pump is equipped with the temperature management system TMS TMH TMU 1600M C T see rating plate heating is possible in the area of the pressu...

Страница 11: ...01 Close the fore vacuum safety valve if available Where Pumping Station Control Unit TCS 180 is involved the valve closes automatically With water cooling Shut off the water supply Shut off the suppl...

Страница 12: ...e vacuum pressure too high Reduce fore vacuum pressure ambient temperature too high Reduce ambient temperature Pump doesn t attain final pressure Pump dirty Bake out pump If seriously contaminated Req...

Страница 13: ...and invoice this work to you if you have not attached this note This also applies where the operator does not have the facilities to carry out the decontamination work Units which are contaminated mi...

Страница 14: ...otation speed max N2 Cooling type water mbar l s 9 15 Ar Cooling type water mbarl s 4 5 6 Gas throughput with TMS operation at nominal rotation speed max1 N2 Cooling type water mbarl s 7 12 Ar Cooling...

Страница 15: ...mensions M8 8x 190 162 2 0 1 2 8 4 5 1 5 7 5 3 5 5 189 9 37 45 4 5 2 8 5 6 5 2 0 331 282 272 228 6 159 295 250 21 257 384 5 397 5 130 DN 200 ISO K DN 250 ISO K DN 200 CF F DN 250 CF F DN 40 ISO KF DN...

Страница 16: ...on 2 1 16 USIT ring 2 P 3529 133 A 54 Venting Valve TVF 005 1 G 1 8 PM Z01 135 see Section 2 1 97 Temperature sensor complete 1 PM 073 593 X see Section 3 9 Spare parts for the air cooling unit 36 Fan...

Страница 17: ...bration compensator TMH DN 200 ISO K PM 006 668 X DN 250 ISO K PM 006 670 X TMU DN 200 CF F PM 006 669 X DN 250 CF F PM 006 671 X Fore Vacuum Safety Valve TVV 001 115 V PM Z01 206 PM 800 263 BN 230 V...

Страница 18: ...he repair and or service of vacuum components will only be carried out if a correctly completed declaration has been submitted Non completion will result in delay The manufacturer could refuse to acce...

Страница 19: ...certify that the product specified below is intended for installation in a machine which is for bidden to be put into operation until such time as it has been determined that the end product is in ac...

Страница 20: ...elefax 662 259 6243 U S A Pfeiffer Vacuum Inc 24 Trafalgar Square Nashua NH 03063 1988 USA telephone 1 603 578 6500 telefax 1 603 578 6550 Venezuela Secotec S A Apartado 3452 Caracas 1010 A telephone...

Отзывы: