6
Product description
3
Product description
3.1 Product identification
This product has been tested to the requirements of CAN/CSA-C22.2 No. 61010-1, sec-
ond edition, including Amendment 1, or a later version of the same standard incorporat-
ing the same level of testing requirements.
For information about other certifications, if applicable, please see the signet on the prod-
uct or:
●
www.tuvdotcom.com
●
TUVdotCOM-ID 0000021320
Product features
The electronic drive unit TM 700 DN is an integrated component of the turbopump. It's
purpose is to drive, monitor and control the entire pump.
To correctly identify the product when communicating with Pfeiffer Vacuum, always have
the information from the rating plate available.
Scope of delivery
●
CD-ROM for DeviceNet connection with EDS files
3.2 Range of application
Pfeiffer Vacuum electronic drive units TM 700 DN must be installed and operated in the
following ambient conditions.
Characteristics
TM 700 DN
Connection voltage TC
48 V DC ± 5 %
Connection panel
DeviceNet
Turbopump HiPace
300 M, 700 M, 800 M
Installation location
weather protected (indoors)
Protection category
IP 54
Protection class
III
Temperature
+5 °C to +40 °C (up to +35 °C with air cooling)
Relative humidity
max. 80 %, at T
≤
31 °C, up to max. 50% at T
≤
40 °C
Atmospheric pressure
75 kPa - 106 kPa
Installation altitude
2000 m max.
Degree of pollution
2
Overvoltage category
II
Содержание TM 700 DN