31
The Pfeiffer Vacuum parameter set
* Only when using pumps with Temperature Management System TMS
Analog output on
"
remote
"
Configuration via parameter
[P:055]
.
Analog input on
"
remote
"
Configuration via parameter
[P:057]
.
Control via interface
Configuration via parameters
[P:060]
and
[P:061]
.
7.4 Operation with the Pfeiffer Vacuum parameter set
Factory settings
The electronic drive unit is pre-programmed in the factory. This guarantees proper, reli-
able turbopump operation without the need for additional configuration.
Checking the adjust-
ments
Before operating with parameters, check set values and control commands for their
suitability for the pumping process.
Remove the remote plug from electronic drive unit if required.
Gas type dependent
operations
Friction causes the rotor to heat up severely under gas load and high rotation speed. To
avoid overheating, the electronic drive unit has implemented power-rotation speed-char-
acteristics, whereby the pump can be operated at every rotation speed with the maxi-
mum allowable gas load without danger of damage. The maximum power consumption
depends on the gas type. Three characteristics are available in order to completely ex-
haust the pump's capacity for each gas type.
12 = Heating
Control is equal to parameter [P:001]
13 = Backing pump
Control is equal to parameter [P:010] and [P:025]
14 = Sealing gas
Control is equal to parameter [P:050]
15 = Pumping station
Control is equal to parameter [P:010]
16 = Pump rotates
active, if rotation speed > 1 Hz
17 = Pump does not rotate
active, if rotation speed < 2 Hz
18 = TMS engaged*
active, if TMS set temperature is engaged
Option
Description
Option
Description
0 = Rotation speed
Rotation speed signal; 0 - 10 V DC = 0 - 100 % x f
Nominal
1 = Power
Power signal; 0 - 10 V DC = 0 - 100 % x P
max
2 = Current
Current signal; 0 - 10 V DC = 0 - 100 % x I
max
3 = always 0 V
always GND
4 = always 10 V
output of continously 10 V DC
5 = follows AI1
follows the analogue input 1
Option
Description
0 = Switched off
Connection is deactivated
1 = Set value in rotation speed setting
mode
Rotation speed setting mode via pin 7 (0 - 10 V) and pin 11
(GND)
Option [P:060]
Description
1 = remote
Operation via connection "
remote
"
2 = RS-485
Operation via connection "
RS-485
"
4 = PV.can
For service purposes only
8 = Field bus
Operation via field bus
16 = E74
Operation via connection "
E74
"
Option [P:061]
Description
0 = off
Interface selection via [P:060]
1 = on
Interface selection locked