Connection "DeviceNet"
Process Components
Instance 2: RS485
101.2.19
Address
USINT
get, set
1-255: RS485 address
Instance 4: Remote
101.4.17
Configuration
USINT
get, set
{0;4}
Instance 6: DI1
101.6.17
Configuration
USINT
get, set
0-5
Instance 7: DI2
101.7.17
Configuration
USINT
get, set
0-5
Instance 8: DI3
101.8.17
Configuration
USINT
get, set
0-5
Instance 9: AI
101.9.17
Configuration
USINT
get, set
0-1
Instance 10: DO1
101.10.17
Configuration
USINT
get, set
0-15
Instance 11: DO2
101.11.17
Configuration
USINT
get, set
0-15
Instance 12: AO1
101.12.17
Configuration
USINT
get, set
0-5
Instance 13: Relay 1
101.13.17
Configuration
USINT
get, set
0-15
Instance 14: Relay 2
101.14.17
Configuration
USINT
get, set
0-15
Instance 15: Relay 3
101.15.17
Configuration
USINT
get, set
0-15
Instance 16: Accessory A1
101.16.17
Configuration
USINT
get, set
0-10
Instance 17: Accessory B1
101.17.17
Configuration
USINT
get, set
0-10
Instance 18: Accessory A2
101.18.17
Configuration
USINT
get, set
0-10
Instance 19: Accessory B2
101.19.17
Configuration
USINT
get, set
0-10
Path
Name
Data type
Service
Comment
102.0.1
Revision
UINT
get
102.0.2
Max Instance
UINT
get
102.0.3
Number of Instances
UINT
get
Instance 1: Backing pump
102.1.17
Configuration
USINT
get, set
0-2
102.1.21
Switch-Off Threshold
UINT
get, set
0-1000 W
102.1.22
Switch-On Threshold
UINT
get, set
0-1000 W
Instance 7: Rotation speed switch point
102.7.17
Configuration
USINT
get, set
0-1
102.7.18
Status
BOOL
get
– 0: Rotation speed switch point
not attained
– 1: Rotation speed switch point
attained
102.7.21
Switchpoint 1
UINT
get, set
50-97 %
102.7.22
Switchpoint 2
UINT
get, set
5-97 %