9
Connection "remote"
5
Connection "
remote
"
Remote control options are provided via the 26-pole D-Sub connector with the designa-
tion ”
remote
“ on the electronic drive unit.
Shielded connectors and cables must be used.
The following information display the factory setting. Configuration is possible
using the Pfeiffer Vacuum parameter set.
5.1 Pin assignment
1
9
10
18
19
26
Pin Function
Designation factory settings
1
+24 V DC output (V+)
Reference voltage for all digital in- and outputs
2
DI1
Enable venting; open: no; V+: yes
3
DI Motor pump
Drive motor; open: off; V+: on
4
DI Pumping station
Open: off; V+: on and error acknowledgement
5
DI Standby
Standby rotation speed; open: off; V+: on
6
DI2
Heating; open: off; V+: on
7
AI+ Rotation speed setting mode Set value in rotation speed setting mode;
2-10 V DC = 20-100% of the nominal rotation speed
8
DO1
Rotation speed switch point attained;
GND:no; V+: yes (I
max
= 50 mA/24 V)
9
DO2
GND: error; V+: no error (I
max
= 50 mA/24 V)
10
DI3
Sealing gas; open: off; V+: on
11
AI- Rotation speed setting mode
GND
Set value in rotation speed setting mode; GND
12
AO1
Actual rotation speed;
0-10 V DC is equivalent to 0-100%; R
L
> 10 k
Ω
13
DI Error acknowledgement
Error acknowledgement: V+ pulse (min 500 ms)
14
DI Remote priority
Control via interface "remote"; open: off
V+: set and priority over other digital inputs
15
Relais 1
Connection to Pin 16 if relay 1 is inactive
16
Relais 1
Rotation speed switchpoint attained;
relay contact 1 (U
max
= 50 V DC; I
max
= 1 A)
17
Relais 1
Connection to Pin 16 if relay 1 is active
18
Relais 2
Connection to Pin 19 if relay 2 is inactive
19
Relais 2
No error; relay contact 2 (U
max
= 50 V DC; I
max
= 1 A)
20
Relais 2
Connection to Pin 19 if relay 2 is active
21
Relais 3
Connection to Pin 22 if relay 3 is inactive
22
Relais 3
Warning; relay contact 3 (U
max
= 50 V DC; I
max
= 1 A)
23
DO Remote priority
GND: off; V+: remote priority active
24
RS-485 D+
according to specifications and Pfeiffer Vacuum protocol
25
RS-485 D-
according to specifications and Pfeiffer Vacuum protocol
26
Ground (GND)
Reference ground for all digital inputs and all outputs