27
The Pfeiffer Vacuum parameter set
036 Cfg Acc B1
Configuration accessory connection B1
0 = Fan (continous operation)
1 = Venting valve, normally closed
2 = Heating
3 = Backing pump
4 = Fan (temperature controlled)
5 = Sealing gas
6 = always 0
7 = always 1
8 = Power failure venting unit
7 RW
0
8
1
x
037 Cfg Acc A2
Configuration accessory connection A2
0 = Fan (continous operation)
1 = Venting valve, normally closed
2 = Heating
3 = Backing pump
4 = Fan (temperature controlled)
5 = Sealing gas
6 = always 0
7 = always 1
8 = Power failure venting unit
7 RW
0
8
3
x
038 Cfg Acc B2
Configuration accessory connection B2
0 = Fan (continous operation)
1 = Venting valve, normally closed
2 = Heating
3 = Backing pump
4 = Fan (temperature controlled)
5 = Sealing gas
6 = always 0
7 = always 1
8 = Power failure venting unit
7 RW
0
8
2
x
045 Cfg Rel R1
Configuration Relay 1
0 = Rot. speed switch point attained
1 = No error
2 = Error
3 = Warning
4 = Error and/or warning
5 = Set speed attained
6 = Pump on
7 = Pump accelerates
8 = Pump decelerates
9 = always 0
10 = always 1
11 = Remote priority active
12 = Heating
13 = Backing pump
14 = Sealing gas
15 = Pumping station
16 = Pump rotates
17 = Pump does not rotate
7 RW
0
17
0
x
046 Cfg Rel R2
Configuration Relay 2
0 = Rot. speed switch point attained
1 = No error
2 = Error
3 = Warning
4 = Error and/or warning
5 = Set speed attained
6 = Pump on
7 = Pump accelerates
8 = Pump decelerates
9 = always 0
10 = always 1
11 = Remote priority active
12 = Heating
13 = Backing pump
14 = Sealing gas
15 = Pumping station
16 = Pump rotates
17 = Pump does not rotate
7 RW
0
17
1
x
#
Display
Designation
Functions
Da
ta
t
y
pe
Ac
ce
ss
Unit
min max
default