The Pfeiffer Vacuum parameter set
7.3 Configuring the connections
The electronic drive unit is pre-configured in the factory. Thereby the turbopump is im-
mediately operational with the necessary functions. The connections of the electronic
drive unit can be configured to suit individual requirements using the parameter set.
Accessory connec-
tion
Configuration via parameters
[P:035]
,
[P:036]
,
[P:037]
or
[P:038]
.
* Only when using pumps with Temperature Management System TMS
Digital inputs on
"
remote
"
Configuration via parameters
[P:062]
,
[P:063]
or
[P:064]
.
Digital outputs and
relays on "
remote
"
Configuration via parameters
[P:019]
and
[P:024]
, respectively
[P:045]
,
[P:046]
,
[P:047]
and
[P:028]
.
●
In the description "active" means:
– For all digital outputs: V+ active high
– For all relays: Contact switch-over according to configuration of
[P:028]
Option
Description
0 = Fan (continous operation)
Control via parameter Pumping station
1 = Venting valve, normally closed
Control via parameter Enable venting, when using a venting
valve which is normally closed.
2 = Heating
Control via parameters Heating and Rotation speed switch-
pont attained
3 = Backing pump
Control via parameters Pumping station and operation
mode backing pump
4 = Fan (temperature controlled)
Control via parameter Pumping station and temperature
thresholds
5 = Sealing gas
Control via parameters Pumping station and Sealing gas
6 = always 0
GND for the control of an external device
7 = always 1
+24 V DC for the control of an external device
8 = Power failure venting unit
Control via parameter Enable venting, when using a power
failure venting unit.
9 = TMS heating unit*
TMS switching box control
10 = TMS cooling unit*
TMS control of cooling water supply
Option
Description
0 = deactivated
Connection deactivated
1 = Enable venting
Control is equal to parameter [P:012]
2 = Heating
Control is equal to parameter [P:001]
3 = Sealing gas
Control is equal to parameter [P:050]
4 = Run-up time control
Control is equal to parameter [P:004]
5 = Rotation speed setting mode
Control is equal to parameter [P:026]
Option
Description
0 = Rotation speed switchpoint attained
active, if switchpoint attained
1 = No error
active, if failure-free operation
2 = Error
active, if error message is active
3 = Warning
active, if warning message is active
4 = Error and / or warning
active, if error and / or warning is active
5 = Set rotation speed attained
active, if set rotation speed is attained
6 = Pump on
active, if Pumping station and Motor is on; No Error
7 = Pump accelerates
active, if Pumping station is on;
Actual rotation speed < Set rotation speed
8 = Pump decelerates
active, if Pumping station is on;
Actual rotation speed > Set rotation speed
Pumping station is off;
Rotation speed > 3 Hz
9 = always 0
GND for the control of an external device
10 = always 1
+24 V DC for the control of an external device
11 = Remote priority active
active, if Remote priority is active