![Pfeiffer Vacuum PrismaPro QMA 250 F1 Скачать руководство пользователя страница 78](http://html1.mh-extra.com/html/pfeiffer-vacuum/prismapro-qma-250-f1/prismapro-qma-250-f1_translation-of-the-original_1534986078.webp)
Parameter
QMG 250 M1
QMG 250 M2
QMG 250 M3
Vacuum required
(max. operating pressure)
< 5 × 10
-4
hPa
Maximum underground cross-sensitiv-
ity at 2 u
< 5 ppm
< 65 ppm
< 100 ppm
Detection limit (EM) with QME 250 (straight)
Open ion source
3.0 × 10
-15
hPa
4.0 × 10
-15
hPa
5.0 × 10
-15
hPa
C/B ion source
3.0 × 10
-13
hPa
6.0 × 10
-13
hPa
1.0 × 10
-12
hPa
Grid ion source
1.0 × 10
-13
hPa
2.0 × 10
-13
hPa
4.0 × 10
-13
hPa
Detection limit (EM) with QME 250 (angled)
Open ion source
1.0 × 10
-14
hPa
2.0 × 10
-14
hPa
4.0 × 10
-14
hPa
C/B ion source
3.0 × 10
-13
hPa
6.0 × 10
-13
hPa
1.2 × 10
-12
hPa
Grid ion source
1.0 × 10
-13
hPa
2.0 × 10
-13
hPa
4.0 × 10
-13
hPa
Sensitivity with QME 250 (straight)
10)
Open ion source (Faraday)
> 5 × 10
-4
A/hPa
> 4 × 10
-4
A/hPa
> 3 × 10
-4
A/hPa
C/B ion source (Faraday)
> 2 × 10
-5
A/hPa
> 1 × 10
-5
A/hPa
> 5 × 10
-6
A/hPa
Grid ion source (Faraday)
> 1 × 10
-4
A/hPa
> 5 × 10
-5
A/hPa
> 3 × 10
-5
A/hPa
Sensitivity with QME 250 (angled)
11)
Open ion source (Faraday)
> 5 × 10
-4
A/hPa
> 4 × 10
-4
A/hPa
> 3 × 10
-4
A/hPa
C/B ion source (Faraday)
> 2 × 10
-5
A/hPa
> 1 × 10
-5
A/hPa
> 5 × 10
-6
A/hPa
Grid ion source (Faraday)
> 1 × 10
-4
A/hPa
> 5 × 10
-5
A/hPa
> 3 × 10
-5
A/hPa
Usage
Installation location
Indoor, weatherproof
Ambient conditions
Protection degree
IP 30
Relative humidity
≤ 98 % at operating temperature, non-condensing
Temperatures
Operating temperature (analyzer)
≤ 150 °C
Operating temperature (QME)
5 to 50 °C
Bakeout temperature (analyzer)
(QME disassembled)
≤ 300 °C
Temperature (transport / storage)
-25 to +70 °C
Temperature coefficient
(during 8 hours, after 30 min. warm-up
time)
< 1 % of the peak height per °C
Electrical data
Supply voltage (power supply pack)
100 to 240 V AC, 50/60 Hz
Connection voltage (QME 250)
20 to 30 V DC, typically 24 V DC
Rated (nominal) current
1.25 A max.
Interfaces
Operation, operating
Ethernet
Standard control connection (AUX I/O)
Power supply plug
D-Sub, 15-pin, female
Relay output
1×, two-way contacts, max. 24 V DC, 0.5 A
Analog inputs
1× 0 to +10 V, output resistance 1 MΩ, resolution 12 bit
Digital inputs
2×; low < +2 V; high > +3 V, n24 V, max. +28 V
Output voltage
+24 V, max. 1 A
Extended control connection (AUX I/O, optional)
10) In EM mode, significantly increased sensitivities can be achieved, depending on the voltage at the EM.
11) In EM mode, significantly increased sensitivities can be achieved, depending on the voltage at the EM.
Technical data and dimensions
78/86
Содержание PrismaPro QMA 250 F1
Страница 1: ...OPERATING INSTRUCTIONS EN Translation of the Original QMG 250 PRISMAPRO Compact mass spectrometer ...
Страница 85: ...85 86 ...
Страница 86: ......
Страница 87: ......
Страница 88: ... BG6001 ed E Date 2101 P N BG6001BEN ...