Parameter
QMG 250 F1
QMG 250 F2
QMG 250 F3
Operating temperature (QME)
5 to 50 °C
Bakeout temperature (analyzer)
(QME disassembled)
≤ 300 °C
Temperature (transport / storage)
-25 to +70 °C
Temperature coefficient
(during 8 hours, after 30 min. warm-up
time)
< 1 % of the peak height per °C
Electrical data
Supply voltage (power supply pack)
100 to 240 V AC, 50/60 Hz
Connection voltage (QME 250)
20 to 30 V DC, typically 24 V DC
Rated (nominal) current
1.25 A max.
Interfaces
Operation, operating
Ethernet
Standard control connection (AUX I/O)
Power supply plug
D-Sub, 15-pin, female
Relay output
1×, two-way contacts, max. 24 V DC, 0.5 A
Analog inputs
1× 0 to +10 V, output resistance 1 MΩ, resolution 12 bit
Digital inputs
2×; low < +2 V; high > +3 V, n24 V, max. +28 V
Output voltage
+24 V, max. 1 A
Extended control connection (AUX I/O, optional)
Power supply plug
HD D-Sub, 62-pin, female
Analog outputs
8× 0 to +10 V, output resistance 100 Ω, I
max.
= 10 mA, resolution 16 bit
Analog inputs
8× -10 to +10 V, input resistance 50 kΩ, resolution 16 bit
Digital outputs
10×; open collector, n24 V, max. +28 V, 200 mA
Digital inputs
5×, low active, low < +2 V; high > +3 V, ≤ +24 V
Output voltage
+24 V, max. 1 A (including current to +24 V output at standard control connection)
Mounting orientation
Venting requirements
min. 25 mm open space around the QME 250
Mounting orientation
Arbitrary
Weight
QMG 250 F analyzer
≤0.8 kg
Electronic unit QME 250
≤1.53 kg
SP 250 Power supply unit
≤0.15 kg
Substances in contact with the media
Filament
Tungsten (W) or Yttrium oxide-coated iridium (Ir-Y
2
O
3
)
Other substances
stainless steel, ceramic (Al
2
O
3
), glass, gold
Tbl. 30:
Technical data PrismaPro QMG 250 F
15.3 Technical data PrismaPro QMG 250 M
Standard conditions
All values were measured after a 30 minute heat-up phase under standard conditions as
long as not specified otherwise.
Parameter
QMG 250 M1
QMG 250 M2
QMG 250 M3
Detector type
EM / Faraday
Mass range
1 to 100 u
1 to 200 u
1 to 300 u
Connection flange
DN 40 CF-F
Technical data and dimensions
77/86