12
Product description
4.2.2
Cooling
●
Air cooling
●
Water cooling (optional)
In the case of excess temperature the electronic drive unit reduces the drive power au-
tomatically.
4.3
Range of application
The pumping station HiCube Pro must be installed and operated in the following ambient
conditions.
Installation location
weather protected (indoors)
Protection category
IP 20
Protection class
I
Temperature
+5 °C to +40 °C (up to +35 °C with air cooling)
Relative humidity
max. 80 %, at T
31 °C, up to max. 50% at T
40 °C
Atmospheric pressure:
750 hPa - 1060 hPa
Installation altitude
2000 m max.
Degree of pollution
2
Overvoltage category
II
Connection voltage TC
24 or 48 V DC ±5%
Limit values for
HiPace 80
HiPace 300
HiPace 400
HiPace 700
Permissible magnetic field max. 3.3 mT
5.5 mT
6 mT
6 mT
Permissible irradiated heat input 3 W
8 W
14 W
14 W
Remarks to ambient conditions
The specified permissible ambient temperatures apply to operation of the turbopump at
maximum permissible fore-vacuum pressure or at maximum gas throughput depending
on the cooling method. The turbopump is intrinsically safe by a redundant temperature
monitoring.
●
By reducing the fore-vacuum pressure or gas throughput, the turbopump can be op-
erated at higher ambient temperatures.
●
If the maximum permissible operating temperature of the turbopump is exceeded, the
electronic drive unit reduces drive power first and switches off then, if necessary.