Tools menu
104
LANCE normalization
Note: LANCE must be installed for this to be enabled. In the enhanced security mode this
tool is enabled only for users who belong to the Editor user group.
Using the LANCE normalization wizard
Before LANCE time-resolved measurements can be made, the instrument must be
normalized for the protocol to be used. This procedure corrects for the following effects in
LANCE measurements: background from the plate, crosstalk of the donor signal into the
acceptor window, and quenching. The final result is normalized to the maximum donor
intensity. To begin this procedure select LANCE normalization from the Wallac 1420
Manager Tools menu. This will activate a wizard that will guide you through the
normalization procedure.
First comes a welcome to the wizard page.
You can select that this does not appear in the future. Click
Next
to proceed. After this the
main steps in the normalization are:
Содержание Wallac 1420
Страница 2: ......
Страница 10: ......
Страница 11: ...9 Introduction...
Страница 12: ...10...
Страница 17: ...15 Operating Wallac 1420...
Страница 18: ...16...
Страница 36: ...Operating Wallac 1420 34...
Страница 37: ...35 Protocol editing...
Страница 38: ...36...
Страница 59: ...57 Result viewing...
Страница 60: ...58...
Страница 67: ...65 Tools menu...
Страница 68: ...66...
Страница 128: ......
Страница 129: ...127 Exiting...
Страница 130: ...128...
Страница 132: ...Exiting 130...
Страница 133: ...131 Troubleshooting...
Страница 134: ...132...
Страница 138: ...Troubleshooting 136...
Страница 139: ...137 Appendices...
Страница 140: ...138...
Страница 148: ...Appendix 3 Dispenser operation 146 Preparing the tubing Put the tubing into a container of water...
Страница 165: ...163 Index...
Страница 166: ...164...