Tools menu
102
Strip A should be full of blank samples, Strip B should be empty and Strip C should be full
of europium samples.
Measurement status page
This is shown during the actual normalization measurement and it gives you feedback about
the measurement. The measurement status pane shows the current status of the
measurement. The progress of the measurement is shown by the progress bar. The
Next
button becomes available when the measurement has been completed. When the
measurement is completed click
Next
to move the wizard directly to the Normalization
confirmation page, otherwise the wizard moves to the Measurement warning page.
Measurement warning page
This is used to inform you that something out of the ordinary has happened during the
measurement of the normalization assay. This page gives you enough information to decide
whether the results can be accepted regardless of the problem that caused the warnings.
Clicking
Next
on this page advances the wizard to the normalization confirmation page.
Clicking
Cancel
closes the wizard and the normalization factors are not recorded in the
database.
Содержание Wallac 1420
Страница 2: ......
Страница 10: ......
Страница 11: ...9 Introduction...
Страница 12: ...10...
Страница 17: ...15 Operating Wallac 1420...
Страница 18: ...16...
Страница 36: ...Operating Wallac 1420 34...
Страница 37: ...35 Protocol editing...
Страница 38: ...36...
Страница 59: ...57 Result viewing...
Страница 60: ...58...
Страница 67: ...65 Tools menu...
Страница 68: ...66...
Страница 128: ......
Страница 129: ...127 Exiting...
Страница 130: ...128...
Страница 132: ...Exiting 130...
Страница 133: ...131 Troubleshooting...
Страница 134: ...132...
Страница 138: ...Troubleshooting 136...
Страница 139: ...137 Appendices...
Страница 140: ...138...
Страница 148: ...Appendix 3 Dispenser operation 146 Preparing the tubing Put the tubing into a container of water...
Страница 165: ...163 Index...
Страница 166: ...164...