Preparing Samples . 61
To thin a sample in the miniature diamond anvil cell:
1.
Loosen and remove the three screws that hold the cell together.
2.
Lift off the top half of the cell and set it aside.
3.
Place the sample on the bottom half of the cell. (The sample must be small.)
4.
Put the top half back on the cell, lining up the red lines on the top and bottom halves.
Do not tighten the screws yet; applying uneven shear forces may damage the
diamonds.
CAUTION
One diamond can damage another.
5.
Press down on the cell with your thumb to thin the sample.
6.
Replace the three screws and tighten them finger tight.
NOTE:
If the spectrum collected with the diamond anvil cell shows interference fringes,
place some KBr in the cell and collect a background spectrum through it.
Figure 18 The Miniature Diamond Anvil Cell
Pressing elastomers between windows
If your sample is elastic and you are compressing it between windows, you must apply
pressure continuously. Use the following procedure:
1.
Press on the windows with a probe, flattening the sample.
2.
While maintaining the pressure, apply small amounts of quick-setting nitrocellulose
cement to the edges of the salt plates.
When the cement is dry, the sample remains compressed.
on page 66) enables you to compress
this kind of sample.
Содержание Spotlight 200
Страница 1: ...Spotlight 200 User s Guide MOLECULAR SPECTROSCOPY...
Страница 6: ...6 Spotlight 200 User s Guide...
Страница 7: ...Introduction...
Страница 13: ...Warnings and Safety Information...
Страница 28: ...28 Spotlight 200 User s Guide...
Страница 29: ...Overview of the Spotlight 200...
Страница 40: ...40 Spotlight 200 User s Guide...
Страница 41: ...Getting Ready to Use the Spotlight 200...
Страница 51: ...Preparing Samples...
Страница 65: ...Techniques for Collecting Spectra...
Страница 112: ...Maintenance...
Страница 126: ...126 Spotlight 200 User s Guide...
Страница 127: ...Appendices...