Maintenance . 123
Connections to the Stage Controller
The connection labeled
X
MOTOR
on the stage controller plugs into the connector next to
the red dot on the motorized stage.
The X, Y and Z connectors are 15-way D-type connectors. They are described below:
Pin
Description
Voltage
Current
1.
OUT1
24 V max
2.52 A peak
2.
OUT2
24 V max
2.52 A peak
3.
OUT3
24 V max
2.52 A peak
4.
OUT4
24 V max
2.52 A peak
5.
X_
END;
Y_
END;
Z_
END;
+5 V max
(Input signal)
6.
X_
ZERO;
Y_
ZERO;
Z_
ZERO
+5 V max
(Input signal)
7.
+5 V
+5 V
2.4 A maximum available
8.
OV
L
OV
L
(2.1 A short circuit)
9.
OUT1
24 V max
2.52 A peak
10.
OUT2
24 V max
2.52 A peak
11.
OUT3
24 V max
2.52 A peak
12.
OUT4
24 V max
2.52 A peak
13.
X_
END;
Y_
END;
Z_
END;
+5 V max
(Input signal)
14.
X_
ZERO;
Y_
ZERO;
Z_
ZERO
+5 V max
(Input signal)
15.
+24 V
+24 V
2.52 A peak
Содержание Spotlight 200
Страница 1: ...Spotlight 200 User s Guide MOLECULAR SPECTROSCOPY...
Страница 6: ...6 Spotlight 200 User s Guide...
Страница 7: ...Introduction...
Страница 13: ...Warnings and Safety Information...
Страница 28: ...28 Spotlight 200 User s Guide...
Страница 29: ...Overview of the Spotlight 200...
Страница 40: ...40 Spotlight 200 User s Guide...
Страница 41: ...Getting Ready to Use the Spotlight 200...
Страница 51: ...Preparing Samples...
Страница 65: ...Techniques for Collecting Spectra...
Страница 112: ...Maintenance...
Страница 126: ...126 Spotlight 200 User s Guide...
Страница 127: ...Appendices...