Preparing Samples . 63
Separating by aperturing
Powders and other particulate solids may contain several different components. Instead of
separating them, use the infrared aperture to isolate the component you want to sample:
1.
Spread the sample out with a probe so that you can visually distinguish the
components.
2.
Looking at the Camera View window, find a particle of the component you want to
sample.
3.
Center this particle in the field of view.
4.
Adjust the infrared aperture (shown by the red dashed lines) until only the particle that
is of interest is visible.
You can easily pick up extremely small particles and transfer them with a very fine-pointed
tungsten needle. Scoring the surface of the salt plate with the needle makes a simple map to
help you positively identify the particles under the microscope.
Transferring with a tungsten needle
When necessary, sharpen the tungsten needle.
Nujol or fluorolube mulls
Suspending fine particles of a solid sample in nujol or fluorolube reduces or eliminates the
surface reflections that can distort absorption measurements. These oils also reduce the
amount of radiation lost to reflection or scattering.
If the film is thin enough, you can correct the spectrum for the presence of the oil by
subtracting a spectrum of the pure liquid. It is difficult, however, to obtain the correct
thickness for a good subtraction.
Fibers
You can roll fibers to flatten them (see
on page 58), or they can be pressed
on page 60).
Fibrous Solids
If a fibrous sample, such as paper, is too thick, tear it and examine the torn edges. The
edges contain single fibers and thin clumps of fibers.
Содержание Spotlight 200
Страница 1: ...Spotlight 200 User s Guide MOLECULAR SPECTROSCOPY...
Страница 6: ...6 Spotlight 200 User s Guide...
Страница 7: ...Introduction...
Страница 13: ...Warnings and Safety Information...
Страница 28: ...28 Spotlight 200 User s Guide...
Страница 29: ...Overview of the Spotlight 200...
Страница 40: ...40 Spotlight 200 User s Guide...
Страница 41: ...Getting Ready to Use the Spotlight 200...
Страница 51: ...Preparing Samples...
Страница 65: ...Techniques for Collecting Spectra...
Страница 112: ...Maintenance...
Страница 126: ...126 Spotlight 200 User s Guide...
Страница 127: ...Appendices...