100 . Spotlight 200 User’s Guide
5.
Click on the Stage View toolbar to collect an image survey.
6.
Ensure that the
ATR
sampling mode is selected in the Setup Microscope Basic tab.
7.
Use
Monitor
to find the most suitable stage height to give a staisfactory spectrum of
your sample.
Refer to
Collecting a Sample Spectrum
on page 98. Click
Halt
when completed, but do
not change the stage height.
8.
Place markers, lines or maps on to your image survey as required.
For information on how to do this, see the on-screen Help. The markers will all be set at
the stage height (Z-axis position) that you determined previously using Monitor.
9.
Click the
Scan Markers
button on the Measurement toolbar.
10.
When prompted, lower the ATR crystal to the working position and click
Retry
.
The stage will move to the position of each marker in turn and collect a spectrum. The
results will be displayed in an Image View.
CAUTION
Always retract the ATR crystal immediately after data collection has
been completed, in order to avoid accidental damage.
Automated ATR Objective
Collecting a Sample Spectrum
1.
Select the
ATR
sampling mode on the Setup Microscope Basic tab.
2.
Choose the
Scan Settings
you want on the Setup Instrument Basic tab.
3.
Move the sample to the center of the Camera View.
Even though the microscope may not be focused on the sample, you can see a change
in the light intensity as the sample moves into the center of the field of view.
4.
Look at the image in the Camera View pane and focus on the sample using the
Z-control
on the joystick, or the
Auto-Focus
option, and then move the area of interest
into the center of the field of view.
5.
Set the infrared aperture to a suitable size and rotation.
For information on how to do this see the on-screen Help.
The contact area of the ATR objective is approximately 100 µm in diameter. We
therefore recommend that you use a maximum aperture size of 100 x 100 µm.
6.
Click
to collect a background spectrum.
The stage moves downwards a small distance and the ATR crystal is lowered to collect
the background spectrum. Once complete, the crystal is raised and the stage returns to
its original position.
Содержание Spotlight 200
Страница 1: ...Spotlight 200 User s Guide MOLECULAR SPECTROSCOPY...
Страница 6: ...6 Spotlight 200 User s Guide...
Страница 7: ...Introduction...
Страница 13: ...Warnings and Safety Information...
Страница 28: ...28 Spotlight 200 User s Guide...
Страница 29: ...Overview of the Spotlight 200...
Страница 40: ...40 Spotlight 200 User s Guide...
Страница 41: ...Getting Ready to Use the Spotlight 200...
Страница 51: ...Preparing Samples...
Страница 65: ...Techniques for Collecting Spectra...
Страница 112: ...Maintenance...
Страница 126: ...126 Spotlight 200 User s Guide...
Страница 127: ...Appendices...