94 . Spotlight 200 User’s Guide
It may be necessary to use a non-marring implement to clean any debris on the sliding
surfaces (I use a finger nail). Take care not to damage the crystal. This is a good time
to clean the crystal itself, refer to the User Guide for suggestions.
10.
Once all of the methanol has evaporated, it’s time to reassemble the system. Insert the
crystal holder into the bore of the bearing being careful to line up the groove with the
ball in the crystal holder.
11.
Gently press down on the crystal holder until it stops. You may need to twist the holder
a small amount from side to side to seat it on the retracted plunger.
Do not force it
,
you may damage the stop contact.
12.
Using another swab, clean any debris from the knurled stop collar. Notice the black
spots in the image below.