Operation
95
Initial (default) value: blank
Upon entering a memorize spectrum No., the previous spectral data will be
deleted and the new data will be memorized.
When the Xe lamp goes off, press the ENT key and the 1st monitor screen will
return.
f.
The scan monitor screen appears, and spectral acquisition within the set range is
carried out.
<Scan monitor screen (spectral acquisition under way)>
SCANNING SPECTRUM NO.: 1 MODE: FL (Em)
Ex: 250 nm Em: *** nm
Pressing ESC key during spectral acquisition will cancel the process. Spectral data
will not be memorized.
The memorize spectrum No. set in the preceding step will be indicated at SCANNING
SPECTRUM NO.
Em is effective within the scan range set above.
g.
When the spectral data memorization is completed, the system returns to the 1st monitor
screen automatically.
Setting of Recorder and Spectrum Output Factors
The following items are settable as factors used in recorder and/or spectrum output.
•
Full scale range of recorder
•
Output speed of recorder
•
Output spectrum data
NOTE:
With an emission spectrum, scattered excitation light, Raman scattering due to solvent, emission due to
solvent or impurities in it, and second (or third) order emission may be superimposed on the emission of
the analyte element. Similarly, an excitation spectrum may be affected by scattered light at the emission
wavelength, by Raman scattering due to solvent, by emission due to solvent or impurities contained in it,
and by second (or third) order scattered light of the excitation beam. These factors must be taken into
consideration when selecting the optimum excitation and emission wavelengths for spectrum acquisition.
Refer to section 3.4 for details.
NOTE:
The excitation and emission spectra output by the Series 200a are not corrected for the wavelength-
dependent sensitivity of the photomultiplier. The obtained spectra therefore differ from the true
excitation and emission spectra of the analyte element.
This difference is especially noticeable at wavelengths below 300 nm.
1.
Full Scale Range of Recorder
a.
Press the RECORD key.
b.
The recorder setting screen appears.
<Recorder setting screen>
Содержание 200A Series
Страница 1: ...SERIES 200A FLUORESCENCE DETECTOR User s Guide LIQUID CHROMATOGRAPHY ...
Страница 5: ...Introduction ...
Страница 6: ......
Страница 15: ...Safety Practices ...
Страница 16: ......
Страница 17: ......
Страница 33: ...Safety Practices 33 Figure 2 2 Overview of the warning labels on the front of the instrument ...
Страница 34: ...34 Series 200a Fi User s Guide Figure 2 3 Detail of the warning labels on the front of the instrument ...
Страница 35: ...System Description 35 System Description ...
Страница 36: ...36 Series 200a Fi User s Guide ...
Страница 55: ...Installation 55 Installation ...
Страница 56: ...56 Series 200a Fi User s Guide ...
Страница 66: ...66 Series 200a Fi User s Guide ...
Страница 67: ...Operation 67 Operation ...
Страница 68: ...68 Series 200a Fi User s Guide ...
Страница 117: ...Maintenance 117 Maintenance ...
Страница 118: ...118 Series 200a Fl User s Guide ...
Страница 148: ...148 Series 200a Fl User s Guide Fig 6 18 The Xenon Lamp Fig 6 19 Orientation of Lamp Bulb Protrusion ...
Страница 154: ...154 Series 200a Fl User s Guide ...
Страница 155: ...Troubleshooting ...
Страница 156: ...156 ...
Страница 162: ...162 Series 200a Fl User s Guide ...
Страница 163: ...Appendix 163 Appendix ...
Страница 164: ...164 Series 200a Fl User s Guide ...
Страница 170: ...170 Series 200a Fl User s Guide ...