Safety Practices .
21
Electric Shock Due to Contact with Inside of Instrument
•
The xenon lamp is supplied with high voltage of 30 kV, which could cause an electric shock
to result in serious or fatal injury. Before removing the light source cover for replacement
of the light source lamp, be sure to turn off the power switch and unplug the power cord
before proceeding.
•
When replacing the fuse with a new one, be careful not to receive an electric shock. To
prevent this, be sure turn off the power switch and unplug the power cord before
proceeding to fuse replacement.
Electric Shock Due to Improper Grounding
To prevent an electric shock hazard, a properly grounded power supply outlet must be provided for
the instrument.
Be sure to use the grounded three prong power cable, which is supplied as a standard accessory
for the instrument. The use of a different type of power cable may result in an electric shock
hazard. Connect the three prong power cable to a grounded three prong power outlet.
Injury Due to Xenon Lamp Explosion
Replace the xenon lamp with a new one before its cumulative turn-on time exceeds the guaranteed
useful lifetime.
With the lapse of the guaranteed useful lifetime, the bulb wall darkens due to evaporation of
electrode material and adhesion of scattered substances, resulting in a decrease in heat dissipation.
This may increase temperature (pressure) inside the bulb excessively to cause
explosion.
Lamp Replacement
Replace the xenon lamp with a new one before its cumulative turn-on time exceeds the guaranteed
useful lifetime.
P/N
Part
Name
Limit
N292-2082
150 W xenon lamp
500 hours
WARNING
Before removing the xenon lamp for replacement, turn off the xenon
lamp (turn off power to the instrument and unplug the power cord)
and then wait for at least one hour until the xenon lamp becomes
sufficiently cool to reduce its internal pressure to the normal safety
level.
WARNING
If a strong shock or impact is applied to the xenon lamp or if the
surface of its quartz glass part is scratched, it may explode and
scatter glass pieces, resulting in personal injury.
Содержание 200A Series
Страница 1: ...SERIES 200A FLUORESCENCE DETECTOR User s Guide LIQUID CHROMATOGRAPHY ...
Страница 5: ...Introduction ...
Страница 6: ......
Страница 15: ...Safety Practices ...
Страница 16: ......
Страница 17: ......
Страница 33: ...Safety Practices 33 Figure 2 2 Overview of the warning labels on the front of the instrument ...
Страница 34: ...34 Series 200a Fi User s Guide Figure 2 3 Detail of the warning labels on the front of the instrument ...
Страница 35: ...System Description 35 System Description ...
Страница 36: ...36 Series 200a Fi User s Guide ...
Страница 55: ...Installation 55 Installation ...
Страница 56: ...56 Series 200a Fi User s Guide ...
Страница 66: ...66 Series 200a Fi User s Guide ...
Страница 67: ...Operation 67 Operation ...
Страница 68: ...68 Series 200a Fi User s Guide ...
Страница 117: ...Maintenance 117 Maintenance ...
Страница 118: ...118 Series 200a Fl User s Guide ...
Страница 148: ...148 Series 200a Fl User s Guide Fig 6 18 The Xenon Lamp Fig 6 19 Orientation of Lamp Bulb Protrusion ...
Страница 154: ...154 Series 200a Fl User s Guide ...
Страница 155: ...Troubleshooting ...
Страница 156: ...156 ...
Страница 162: ...162 Series 200a Fl User s Guide ...
Страница 163: ...Appendix 163 Appendix ...
Страница 164: ...164 Series 200a Fl User s Guide ...
Страница 170: ...170 Series 200a Fl User s Guide ...