92.
Series 200a Fi User’s Guide
NOTE:
The spectral bandwidth is 15 nm on both excitation and emission sides with the Series 200a FL detector.
A component with a difference between excitation and emission wavelengths of less than 30 nm cannot
be accurately measured. Excitation and emission wavelengths should be kept more than 30 nm apart for
measurement.
NOTE:
The maximum wavelength scan range (from end to start wavelength) for both excitation and emission is
400 nm.
For spectrum acquisition in a range greater than
400 nm, the range should be divided into two for measurement.
NOTE:
Saved spectral data will not be retained when the power is turned OFF.
1.
Measurement and Recording of Spectrum for Excitation (Ex) Scan Range:
a.
Press the SPECT key.
b.
The scan mode setting screen will appear.
<Scan mode setting screen>
SPECTRUM MODE (FL (Ex)=1, FL (Em)=2)
1
Press numeric key 1 and then ENT to select FL (Ex).
Now proceed to the scan range setting screen for setting a scan range, and enter a start
wavelength (cursor will appear at WL1).
c.
Enter a start wavelength (minimum value) at WL1 of the scan range.
<Scan range setting screen (indicate WL1)>
SCANNING WL RANG (WL1 WL2) nm FL (Ex)
<200-800> 200 - 600
Press numeric keys 2 0 0 to 8 0 0 and then ENT.
Setting range: 200 to 800 nm
Initial (default) value: 200
If setting will not be changed, press
V
key to proceed to WL2 setting.
The cursor now moves to WL2.
d.
Now enter an end wavelength (maximum value) at WL2.
<Scan range setting screen (indicate WL2)>
SCANNING WL RANG
(WL1 WL2) nm FL (Ex)
<250-850>
200 - 600
Содержание 200A Series
Страница 1: ...SERIES 200A FLUORESCENCE DETECTOR User s Guide LIQUID CHROMATOGRAPHY ...
Страница 5: ...Introduction ...
Страница 6: ......
Страница 15: ...Safety Practices ...
Страница 16: ......
Страница 17: ......
Страница 33: ...Safety Practices 33 Figure 2 2 Overview of the warning labels on the front of the instrument ...
Страница 34: ...34 Series 200a Fi User s Guide Figure 2 3 Detail of the warning labels on the front of the instrument ...
Страница 35: ...System Description 35 System Description ...
Страница 36: ...36 Series 200a Fi User s Guide ...
Страница 55: ...Installation 55 Installation ...
Страница 56: ...56 Series 200a Fi User s Guide ...
Страница 66: ...66 Series 200a Fi User s Guide ...
Страница 67: ...Operation 67 Operation ...
Страница 68: ...68 Series 200a Fi User s Guide ...
Страница 117: ...Maintenance 117 Maintenance ...
Страница 118: ...118 Series 200a Fl User s Guide ...
Страница 148: ...148 Series 200a Fl User s Guide Fig 6 18 The Xenon Lamp Fig 6 19 Orientation of Lamp Bulb Protrusion ...
Страница 154: ...154 Series 200a Fl User s Guide ...
Страница 155: ...Troubleshooting ...
Страница 156: ...156 ...
Страница 162: ...162 Series 200a Fl User s Guide ...
Страница 163: ...Appendix 163 Appendix ...
Страница 164: ...164 Series 200a Fl User s Guide ...
Страница 170: ...170 Series 200a Fl User s Guide ...