50.
Series 200a Fi User’s Guide
3.
Second-order and Third-order Radiation
A diffraction grating that is set to transmit radiation of a specific wavelength (e.g. 250 nm) will also
transmit multiples of the radiation (e.g. 500 nm). When a fluorescence spectrum is collected, the
analyst should be aware of the second order (and third order) phenomena and take care that peaks
are not misidentified.
For instance, if the excitation wavelength is 240 nm, the second-order and third-order excitation occur
at 480 and
720 nm, respectively. To eliminate second and third order radiation, it is simply necessary to place a
short-wavelength cutoff filter in the path of fluorescing radiation (before the fluorescence
monochromator).
4.
An Example of a Fluorescence Spectrum
Figure 3-11 provides an example of a fluorescence spectrum, which contains the various types of
peaks indicated above. In most cases, the fluorescence of the compound of interest lies closer to the
excitation than shown in Fig. 3-10, and overlapping of peak of interest and the second order radiation
is rarely a problem. Typically, the peak for the compound of interest might correspond to peak *D,
and the analyst might have to correct for the Raman peak (e.g. use a shorter excitation wavelength).
*A Scattering of excitation beam
*B Raman scattering due to eluent
*C Fluorescence of impurity
*D Fluorescence of compound of interest
*E Second-order scattering of excitation beam
Fig. 3-11 Measurement Example of Fluorescence Spectrum
5.
Temperature Dependency of Fluorescence Intensity
In most cases, the fluorescence intensity decreases by 1 to 2% as the temperature of sample
increases by 1
°
C. It should be noted, however, that in some biochemical samples, the
fluorescence intensity decreases by approx. 10% as the temperature of sample increases by 1
°
C
(especially if the fluorescence is due to the indole ring system).
Relative intensity
Wavelength
EX
×
2
EX
*A
*B
*C
*D
*E
Содержание 200A Series
Страница 1: ...SERIES 200A FLUORESCENCE DETECTOR User s Guide LIQUID CHROMATOGRAPHY ...
Страница 5: ...Introduction ...
Страница 6: ......
Страница 15: ...Safety Practices ...
Страница 16: ......
Страница 17: ......
Страница 33: ...Safety Practices 33 Figure 2 2 Overview of the warning labels on the front of the instrument ...
Страница 34: ...34 Series 200a Fi User s Guide Figure 2 3 Detail of the warning labels on the front of the instrument ...
Страница 35: ...System Description 35 System Description ...
Страница 36: ...36 Series 200a Fi User s Guide ...
Страница 55: ...Installation 55 Installation ...
Страница 56: ...56 Series 200a Fi User s Guide ...
Страница 66: ...66 Series 200a Fi User s Guide ...
Страница 67: ...Operation 67 Operation ...
Страница 68: ...68 Series 200a Fi User s Guide ...
Страница 117: ...Maintenance 117 Maintenance ...
Страница 118: ...118 Series 200a Fl User s Guide ...
Страница 148: ...148 Series 200a Fl User s Guide Fig 6 18 The Xenon Lamp Fig 6 19 Orientation of Lamp Bulb Protrusion ...
Страница 154: ...154 Series 200a Fl User s Guide ...
Страница 155: ...Troubleshooting ...
Страница 156: ...156 ...
Страница 162: ...162 Series 200a Fl User s Guide ...
Страница 163: ...Appendix 163 Appendix ...
Страница 164: ...164 Series 200a Fl User s Guide ...
Страница 170: ...170 Series 200a Fl User s Guide ...